5. Use a T20 Torxhead screwdriver to remove the two screws holding the power supply
system to the top of the back panel.
Figure 486. Removing the Power Module Screws
Screws
6. Grasp the power supply system and lift it out of the instrument. It fits snugly, so you may
have to gently shake it as you lift it out.
7. Disconnect the CI reagent gas supply tubes from the CI reagent gas flow module.
8. Remove the three screws holding the left hand sub panel to the chassis and pull the panel
away to expose the screws that hold the CI reagent gas flow module to the chassis.