22. Pull the tubing out of the manifold. Replace if necessary.
23. Disconnect the collision gas module cable from the distribution board.
Figure 441. Disconnecting the Collision Gas Module Cable from the Distribution Board
24. Disconnect the 1/16 in. tubing from the bulkhead compression fitting located on the left
hand sub panel.
Figure 442. Disconnecting the Cable to the Bulkhead Fitting
Bulkhead
Fitting
25. Use a T20 Torxhead screwdriver to remove the two screws on the collision gas module.
26. Remove the collision gas module and replace if necessary.
Collision Gas Module
Cable Connection
Collision Gas Module
Tubing
3
Advanced Troubleshooting
Replacing the Collision Gas Components
Electrical
Cable
TSQ 9610 Hardware Manual
401