Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
48
TSQ 9610 Hardware Manual
c. (NeverVent AEI Source) With your hand around the stainless steel barrel, twist the
tool until the alignment groove at the end of the barrel aligns with the left notch
around the slot vacated by the vacuum interlock knob.
Figure 36. Inserting the AEI Source Exchange Tool
Left Notch on
VPI
d. (NeverVent AEI Source) Twist and push the stainless steel barrel to the right until it
clicks into place under the notch at the top of the slot.
Figure 37. Twisting the AEI Source Exchange Tool Into Place
Top Notch on
VPI