Thermo Scientific TSQ 9610 Hardware Manual page 369

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
Figure 374. Opening the Manifold Door
11. Open the manifold door and slide the filament board and wires off the pins on the source
interface board.
Figure 375. Disconnecting the Filament Wires
Filament
Filament
Wires
12. Remove the bottom left screw holding the metal heat shield to the manifold.
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
Screws
TSQ 9610 Hardware Manual
339

Advertisement

Table of Contents
loading

Table of Contents