Replacing Components of the Vacuum Interlock

Figure 270. Replaceable Components of the Vacuum Interlock
Valve Axle
Clip
Thermo Scientific
If the components of your vacuum interlock get damaged, you can use the illustration below
to visually locate a component and then follow the process to replace it.
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
(Metal valve body is
shown for illustration
purposes only. It is not
available for purchase.)
Axle Seal
Axle Bearing
Outer Ball Seal
To replace the components of the vacuum interlock
1. Shut down your system by following the steps in
on
page
139.
Vacuum Interlock Knob
Spring
O-Ring
Inner Ball Seal
Sealing Ball
"Shutting Down the TSQ 9610 System"
3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
TSQ 9610 Hardware Manual
O-Ring
251