Left Hand Front Panel And Left Hand Sub Panel - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Covers of the TSQ 9610 Instrument

Left Hand Front Panel and Left Hand Sub Panel

306
TSQ 9610 Hardware Manual
To replace the left hand front panel and the left hand sub panel of the TSQ 9610
instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139. You do not have to remove the ion source cartridge or column or vent the
system. If you are using hydrogen as a carrier gas, power off the GC.
2. Remove the four screws on the left hand front panel. One is located on the bottom left of
the interior front panel of the instrument.
Figure 341. Locating the Left Hand Front Panel Screws
3. To replace the left hand sub panel, disconnect the inputs as shown in
"Shutting Down the TSQ 9610 System"
Screws
Screw on Front
Figure
342.
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents