Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
342
TSQ 9610 Hardware Manual
16. From the left side of the instrument, reattach the three screws around the 20-pin
feedthrough. Tighten the screws uniformly.
Figure 380. Reattaching the 20-Pin Feedthrough
17. Look through the top of the instrument and align the pins on the 20-pin feedthrough
with the holes in the source interface board to connect them together.
18. Slide the heat shield into its slot and reattach the screw holding it in place.
19. Reattach the filament board to the source interface board.
20. Close the manifold door and reattach the four screws.
20-Pin Feedthrough
Screws
Thermo Scientific