Advanced Troubleshooting
Replacing Components of the Ion Source
236
TSQ 9610 Hardware Manual
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door. See
Figure
244.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
Figure 244. Removing the Manifold Door Screws
9. Open the manifold door. You can now see the ion source block.
Figure 245. Locating the Ion Source Block
10. Insert the small source removal tool into the ion source block as shown in
Ion Source Block
Figure
246.