Thermo Scientific TSQ 9610 Hardware Manual page 351

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
8. Use a 2.5 mm nut driver to remove the standoffs on the rod driver board.
Figure 355. Disconnecting the Rod Driver Board from the RF Board
Screws
9. Incrementally, remove the 40-pin connector on the rod driver board from the front RF
board. and the 7-pin connector on the rod driver board away from the front RF board.
Figure 356. Disconnecting the Rod Driver Board 40-Pin Connector
3
Replacing the Boards in the TSQ 9610 Instrument
Standoffs
Rod Driver Board
40-Pin Connector
TSQ 9610 Hardware Manual
Advanced Troubleshooting
321