Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
64
TSQ 9610 Hardware Manual
3. Align the long tooth of lens 3/RF lens with the notch on the bottom of the sleeve and
drop the lens into the sleeve.
Figure 61. Inserting Lens 3/RF Lens into the Source Sleeve
4. Place lens 2 on top of lens 3/RF lens with the small hole facing down. It should fit snugly
and sit evenly on top of lens 3/RF lens.
Figure 62. Inserting Lens 2 into the Source Sleeve
Long Tooth on Lens 3/RF Lens
Tooth Fits Here
Small Hole of
Lens 2