Thermo Scientific TSQ 9610 Hardware Manual page 313

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
5. Pull the right side panel away from the chassis.
6. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 9610 Instrument"
7. Disconnect the power cable from the left side of the TSQ 9610 instrument.
Figure 319. Disconnecting the Power Cables
Power
Cable
8. Use a T20 Torxhead screwdriver to remove the two screws holding the power supply
system to the top of the back panel.
9. Grasp the power supply system and lift it out of the instrument. It fits snug, so you may
have to shake it as you lift it out.
10. Use a T20 Torxhead screwdriver to remove the four screws holding the back manifold
plate onto the instrument.
3
Advanced Troubleshooting
Replacing Components of the Manifold
on
page
141.
TSQ 9610 Hardware Manual
283