Thermo Scientific TSQ 9610 Hardware Manual page 109

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
16. Pull the handle it back toward you and down the right side of the barrel's metal track until
the line at the end of the track appears.
Figure 85. Pulling the Tool Down the Right Track
17. Pull the vacuum interlock handle down to close the interior ball valve.
Figure 86. Pulling Down the Vacuum Interlock Handle
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
TSQ 9610 Hardware Manual
79

Advertisement

Table of Contents