Advanced Troubleshooting
Replacing Components of the Analyzer
180
TSQ 9610 Hardware Manual
Figure 167. Replacing the Collision Cell and Quadrupole 3 in the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
7. Reattach the wire to the Q3 exit lens.
8. Replace the analyzer tray into the instrument by following the instructions in
the Analyzer Tray in the TSQ 9610 Instrument"
9. Restart your system by following the instructions in
Q3 Exit Lens
Assembly
on
page
154.
"Restarting the System"
Screws
"Replacing
on
page
158.