17. Remove the o-ring and replace it, if necessary.
18. Flip the block over in your hand so that sealing ball and outer ball seal will fall out. Set
them aside or replace them, if necessary.
CAUTION INSTRUMENT DAMAGE: The sealing ball must not be scratched, so be very
careful when handling it and keep it on a clean, lint-free surface.
Figure 279. Replacing the Sealing Ball and Outer Ball Seal
Sealing Ball
Outer Ball Seal
19. Pry the thin metal clip out of the groove. It holds all of the other components inside the
metal valve body.
Figure 280. Replacing the Metal Clip
Clip
3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
TSQ 9610 Hardware Manual
257