21. Reinsert the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 381. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
22. Push the board down into the connector.
23. Look through the left side of the instrument and tighten the screw that holds the lens
driver board in place on the chassis.
24. Reattach the top cover panel and tighten the screw.
25. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
"Restarting the System"
TSQ 9610 Hardware Manual
on
page
158.