21. Close the front door of the instrument.
22. Power-on the TSQ 9610 system.
23. Purge the foreline pump for 10 minutes by using the gas ballast to remove excess gases
from the new oil. See
information.
Purging Gas from the Oil in the Foreline Pump
1
Performing Routine Maintenance
Maintaining the Foreline Pump
for more
TSQ 9610 Hardware Manual