Advanced Troubleshooting
Replacing Fans and Filters
422
TSQ 9610 Hardware Manual
20. Reconnect the turbo cooling fan cable to the distribution board.
21. Replace the fan plenum, chassis cooling fan, and air deflector assembly into the
instrument.
22. Reattach the six screws securing the fan plenum to the chassis.
23. Reconnect the cable on the chassis cooling fan to the distribution board.
24. Replace the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
25. Push the lens driver board down into the connector.
Figure 462. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
26. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
tighten the screw holding the lens driver board in place on the chassis.
27. Replace the right side panel and reattach the four screws.
28. Replace the left hand front panel and reattach the four screws.
29. Replace the top cover panel and reattach the top cover panel screw.
30. Reattach the GC.
31. Restart your system by following the instructions in
"Restarting the System"
on
Thermo Scientific
page
158.