Air Deflector - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

Air Deflector

Thermo Scientific
To replace the air deflector
Note See the ISQ and TSQ GC-MS Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
on
page
139. You do not need to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 452. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
5. Remove the two rivets holding the air deflector to the chassis cooling fan.
3
"Shutting Down the TSQ 9610 System"
Screws
TSQ 9610 Hardware Manual
Advanced Troubleshooting
Replacing Fans and Filters
Screw on Front
of Instrument
413