Section 2: Performance verification
From the front panel:
1. Press the MENU key.
2. Under System, select Calibration.
3. Select Scheduling Action. To have the instrument:
Prompt you to run auto calibration: Select Notify.
Run auto calibration at a specific time: Select Run.
To stop scheduling: Select None. If you select None, you do not need to make additional
settings.
4. Select Scheduling Interval.
5. Select the interval.
6. Select Scheduled Time to select the time when the auto calibration will run or when you will be
prompted to run it.
To review the next schedule time and date, see the information listed next to Next Run.
Using SCPI commands:
Refer to :ACAL:SCHedule.
Using TSP commands:
Refer to
acal.schedule()

Reviewing calibration information

The Calibration screen displays information about the last autocalibration and factory calibrations that
were run and the present status. For detail on this screen, refer to System Calibration menu.
For autocalibration, you can also access this information from the commands in the SCPI ACAL
subsystem or the TSP acal.* commands.
2-6
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
(on page 4-7).
DMM7510-905-01 Rev. A / April 2015