Measuring 4-wire resistance with offset compensation
In this section:
Introduction .............................................................................. 6-1
Equipment required .................................................................. 6-1
Device connections .................................................................. 6-2
4-wire resistance measurements with offset compensation ..... 6-3
Introduction
This application example demonstrates how to use the Model DMM7510 to accurately measure a
resistance device.
To provide the best resistance measurement accuracy, the four-wire (Kelvin) measurement and the
offset compensation methods are used for this test. Offset compensation is a measuring technique
that reduces or eliminates thermoelectric EMFs in low-level resistance measurements. The voltage
offsets because of the presence of thermoelectric EMFs (V
measurement accuracy.
To overcome these offset voltages, you can use offset-compensated ohms.
Equipment required
•
One Model DMM7510
•
Four insulated banana cables, such as the Model 1756 Standard Test Lead Kit that is provided
with the Model DMM7510 (you will need another set)
•
One device to be tested; the application shown here uses a 20 Ω resistor (actual resistance value
can vary with availability and user discretion)
•
One computer set up for communication with the Model DMM7510
北京海洋兴业科技股份有限公司(证券代码:839145)
) can adversely affect resistance
EMF
电话:010-62176775
Section 6
网址:www.hyxyyq.com