Measuring 4-Wire Resistance With Offset Compensation; Introduction; Equipment Required - Keithley DMM7510 User Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Measuring 4-wire resistance with offset compensation

In this section:

Introduction .............................................................................. 6-1

Equipment required .................................................................. 6-1

Device connections .................................................................. 6-2
4-wire resistance measurements with offset compensation ..... 6-3
Introduction
This application example demonstrates how to use the Model DMM7510 to accurately measure a
resistance device.
To provide the best resistance measurement accuracy, the four-wire (Kelvin) measurement and the
offset compensation methods are used for this test. Offset compensation is a measuring technique
that reduces or eliminates thermoelectric EMFs in low-level resistance measurements. The voltage
offsets because of the presence of thermoelectric EMFs (V
measurement accuracy.
To overcome these offset voltages, you can use offset-compensated ohms.
Equipment required
One Model DMM7510
Four insulated banana cables, such as the Model 1756 Standard Test Lead Kit that is provided
with the Model DMM7510 (you will need another set)
One device to be tested; the application shown here uses a 20 Ω resistor (actual resistance value
can vary with availability and user discretion)
One computer set up for communication with the Model DMM7510
北京海洋兴业科技股份有限公司(证券代码:839145)
) can adversely affect resistance
EMF
电话:010-62176775
Section 6
网址:www.hyxyyq.com

Hide quick links:

Advertisement

Table of Contents
loading