2. Loosen the 2 sensor position locking screws.
3. Move the sensor sideways by turning the sensor position adjustment screws (2 pcs).
4. Check that both ends of the sensor have moved evenly from the sensor alignment
holes. The holes are equally visible when the sensor is vertical.
5. Re-tighten the position locking screws after the adjustment.
6. Redo the failed calibration to verify the adjustment.
7. Re-attach the 2D sensor covers.

9.11.6 Patient positioning light adjustment

If the positioning lights are aligned improperly, adjustment is needed to ensure correct
patient positioning.