9.7 Device cover installation

9.7.1 Main device covers

NOTICE! The screws used to attach the covers are located in the screw kit.
1. Detach the device mirror.
2. Attach upper shelf covers. Place the right cover onto the upper shelf and then push
the left cover against it.
3. Secure the covers with 2 screws on the top of the covers and then one on the
bottom.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
141
D519207, 19