9 Installation
5. Slide the wall bracket up to the insertion points on the column and push the wall
bracket into the rail. The rail has openings on both sides at the wall bracket insertion
point.
6. Push the rail cover back to the column.
7. Slide the wall bracket to the required height.
NOTICE! The lower wall bracket installation screws have to be fixed to the
wall 57-87 mm (2.2-3.4") from the floor. See chapter
page 203.
8. Secure the wall bracket in place with 4 pin screws.
9. Re-install the wall bracket rail cover back in place.
10. Re-install the cable clamp to its place.
11. Continue to install the column normally according to chapter
page 87.

9.10.7 Exhibition stand installation

If the device cannot be bolted to the floor, it may be installed on a special exhibition
stand (ordered separately).
NOTICE! For other than exhibition use, the device has to be fixed to the wall
even if the exhibition stand is used.
1. Transport the exhibition stand to the location, where the device is to be installed.
170
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Device dimensions
on
Column installation
on
ORTHOPANTOMOGRAPH
D519207, 19
OP 3D