9 Installation
If a remote exposure switch needs to be used, install it as follows:
1. Connect one end of the remote exposure switch cable to the connector J2 inside the
remote exposure switch.
2. Attach the remote exposure switch to the wall. The wall attachment requires 2 sets
of 3 mm (1/8") diameter fixing hardware.
NOTICE! Guide the remote exposure switch cable through the small cut-
out in the bottom of the remote exposure switch housing and ensure that the
cable does not press against the circuit board.
3. Guide the remote exposure switch cable to the back panel of the device, see chapter
Back panel connectors
4. If the user wants, the exposure button can be attached also to the remote exposure
switch instead of attaching it directly to the device.
168
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
1. Exposure switch
2. Exposure warning light
3. Exposure button connector
on page 86.