9 Installation
3. Re-install the detached covers.
NOTICE! When re-installing the PAN/3D tubehead rear cover you must first
remove the tubehead plate from the PAN/3D tubehead rear cover.
9.11.4 2D/3D Sensor height adjustment
If the device geometry calibration has failed due to the sensor being too high/low the
height adjustment is required.
NOTICE! This adjustment is required only in special cases.
1. Pull the left and right 2D/3D sensor collar cover halves downwards.
2. Pull the right side collar cover half downwards to unlock the teeth holding the covers
together.
3. Pull the collar cover halves apart.