26. Adjust the tilt of the column if needed.
a) Loosen the wall bracket angle locking screws (2 pcs) and adjustment block side
screws (2 pcs).
b) Adjust the column tilt by screwing both tilt adjustment screws simultaneously.
Tilt the column forwards
c) Re-tighten the wall bracket angle locking screws (2 pcs) and adjustment block
side screws (2 pcs).
27. Fix the column to the floor using appropriate screws.
28. Remove the bubble wrap between the outer and inner column covers. Ensure that
the outer column cover plate is properly in place and doesn't scratch the inner
column cover.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
BASIC INSTALLATION
OP 3D
Tilt the column backwards
PIVOT SCREW INSTALLATION
9 Installation
95
D519207, 19