Removing The Patient Handle; Removing The Mirror - KaVo ORTHOPANTOMOGRAPH OP 3D Installation Manual

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

3   Device upgrade
3.4.2   Removing the patient handle
Required tools: Hex key 6mm
1.
Remove the screws (4 pcs) under the patient handle.
2.
Remove the handle.
3.4.3   Removing the mirror
1.
Open the mirror.
2.
Lift the mirror slightly so that the guiding pins (4 pcs) unlock from the keyholes on the mirror plate.
3.
Remove the mirror from its guiding pins by pulling backwards.
NOTICE! Be careful not to damage the mirror when you remove it.