8. Re-attach the mirror by guiding the pins on the carriage plate (4pcs) through the
keyholes on the mirror plate and sliding the mirror down to lock it in place.
9. Route the mains cord, exposure button cable, Ethernet cable and other optional
cables through the cable clamp on the rear of the device.
NOTICE! The left cable clamp screw may need to be loosened to help routing
the thick mains cord into the clamp.
10. Attach the back panel cover by sliding it to its place and secure it with one screw,
located on the left side of the unit.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
143
D519207, 19