26. Push the metallic part of the tubehead cable into the previously attached grounding
clamp.
27. Connect the pre-routed head support cables to the upper connector on the rear of
the device.
28. Attach the cable cover plate with a screw.
29. Once the cephalometric unit has been installed, if needed, adjust the device position
by rotating the whole device by the CEPH arm. See the space requirements for
different positions of the cephalometric device from chapter
dimensions
NOTICE! The device may be rotated when it has been fixed to the floor only
with one screw (column floor plate pivoting screw).
NOTICE! You may need to loosen the wall bracket angle adjustment screws a
bit to enable the device rotation.
NOTICE! Ensure that there's enough room between the CEPH tubehead and
the walls so that the covers can be still installed.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 206.
OP 3D
9 Installation
Cephalometric device
D519207, 19