9.11.5.2 2D Sensor height adjustment
NOTICE! This adjustment is required only in special cases and shouldn't be
needed to be done during the installation of the device.
1. Detach the 2D sensor covers according to chapter
page 182.
2. Loosen the 2 sensor height locking screws.
3. Turn the adjustment screw to adjust the height of the sensor. One turn is equivalent
of 0.7 mm.
4. Re-tighten the height locking screws after the adjustment.
5. Redo the failed calibration to verify the adjustment.
6. Re-attach the 2D sensor covers.