b. Adjust the height of the collimator plate by turning the height adjustment screw;
• Clockwise to move the collimator downwards.
• Counter clockwise to move the collimator upwards.
NOTICE! The GUI informs you how much the collimator needs to be adjusted
and in which direction. One full turn of the screw (360°) equals about 0.7 mm.
To tilt the collimator LEFT/RIGHT
a. Loosen the upper collimator plate adjustment screw.
b. Adjust the tilt of the collimator plate by turning the rotation adjustment screw;
• Clockwise to turn the collimator left.
• Counter clockwise to turn the collimator right.
2. After the adjustments, tighten the locking screws and redo the collimator calibration
to verify the adjustment.