9.11.6.1 Horizontal and midsagittal lights
1. Detach the carriage back plate, back panel cover, mirror, lower shelf tray and
the carriage side covers. See chapter
information.
2. The positioning lights are held in place in a fixture with one screw. Loosen the screw
to release the light.
3. Re-align the light by rotating it and tilting it up/down.
4. Tighten the fixture screw.
5. Adjust the angle of the midsagittal light fixture using the black eccentric screw
located on the right side of the light fixture.
6. Attach the Geometry Calibration Phantom to the device.
7. GUI: Select Device geometry calibration or PAN geometry calibration from
Calibrations menu.
8. All PAN/3D patient positioning lights come on. Check that the Horizontal lights hit the
horizontal grooves and the Midsagittal light hits the vertical groove on the phantom
to ensure that the positioning lights are correctly aligned.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Horizontal lights
OP 3D
Main device covers
on page 141 for more
Midsagittal light
9 Installation
187
D519207, 19