4. Press and hold the exposure button down to take the calibration image.
5. The calibration results appear on the GUI.
NOTICE! This will take several minutes.
6. Acknowledge the calibration by pressing the OK button.
7. Repeat the calibration for the remaining FOV sizes.
7.2.3.4 PAN pixel calibration (2D sensor)
This program calibrates the 2D sensor for Panoramic imaging.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Select PAN pixel calibration (2D sensor) from the Calibrations menu.
2. Protect yourself from radiation.
3. Press and hold the exposure button down to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.