9 Installation

9.8 Acceptance tests

After the device has been completely calibrated and the device covers attached, perform
the device Quality Control programs and store the resulting images for the user as
reference.
A complete list of Quality control programs, required to be performed after the device
installation as part of the acceptance testing can be accessed in Installation mode,
unlocked with PIN code 0612 through the Service mode menu.

9.8.1 PAN QC (optional)

NOTICE! The PAN QC is an optional, yet recommendable, procedure and is
mandatory to be performed where local regulations require it.
NOTICE! The device's country selection needs to be configured per local
requirements before performing the QC programs. See chapter
page 158.
1. GUI: Select PAN QC program from the Quality Control menu.
2. Attach the PAN QC phantom holder and 2D QC test phantom to the device.