9 Installation
7. The calibration image appears on the GUI.
8. Acknowledge the calibration result by pressing the OK button.
9.6.21 CEPH position calibration
This program is used to calibrate the device movements in cephalometric imaging. This
includes the pivot movement of the upper shelf and the linear movement of the rotating
unit.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
CEPH calibration rod is required when performing this calibration.
1. Pull the lid on the center of the CEPH head support down to reveal the calibration rod
insert.
2. Screw the ceph calibration rod in by hand.
3. Turn the CEPH head support to PA position.