10 Technical data
Device data
Model:
Protection against electric shock
Degree of protection
Protection against the ingress of liquids
Cleaning methods
Use environment
Mode of operation
Safety
Source of power
EMC Classification
Tubehead assembly
Tubehead assembly
Tube type
Tube voltage
Max. tube current
Target angle
Focal spot
Total filtration
Additional filtration of cover materials
(CEPH only)
Leakage technique factors
NOTICE! Leakage technique factors may be impacted by ambient temperature or
used workstation.
196
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
PCX–1
Class I
Type B applied with no conductive
connection to the patient
IP20
• Distilled water
• Ethanol 96%
• Isopropyl alcohol
• Soapy water
• Cavicide
disinfectant
• Metasys
disinfectant
• Disposable plastic covers for bite block,
chin rest, lip support and head support
• Disposable plastic covers for ear rods
and nasion support (CEPH only)
In environments where no flammable
anesthetics nor flammable cleaning agents
are present
Continuous operation/intermittent loading
IEC60601–1
Mains connection
Class B
THA/HVGEN PCX
D–054S or equivalent
Stationary anode
60 – 95 kV
16 mA
0.5 (IEC 60336/2005)
min. 3.2 mm Al @ 95kV
min 1.4 mm Al @ 95kV
PAN
4613 mAs/h @ 90kV/12.5mA
3D
2880 mAs/h @ 95kV/4mA
CEPH
4736 mAs/h @ 95kV/11mA
ORTHOPANTOMOGRAPH
OP 3D
D519207, 19