9 Installation
Calibration program order per device configuration:
Device geometry calibration
Collimator calibration
PAN pixel calibration (3D sensor) **
3D geometry calibrations
PAN beam alignment calibration
PAN geometry calibration **
PAN pixel calibration (2D sensor) **
* CEPH device related programs are available only if CEPH is installed.
** Availability of these calibrations depend on which sensor type is used to take PAN
images.
118
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
PAN/3D and
PAN/CEPH/3D
Device movement calibration
X-ray generator calibration
3D pixel calibration
CEPH head support calibration *
CEPH generator calibration *
CEPH tube sweep calibrations *
CEPH secondary collimator calibration *
CEPH geometry calibration *
CEPH pixel calibration *
CEPH position calibration *
CEPH mechanical calibration *
Z-movement limits
PAN geometry calibration
PAN collimator calibration
PAN pixel calibration (2D sensor)
PAN and
PAN/CEPH
ORTHOPANTOMOGRAPH
D519207, 19
OP 3D