9 Installation
9. Re-align the lights if they are still not aligned correctly.
10. Re-install the detached covers.
NOTICE! When attaching the left side cover, ensure that the horizontal lasers
sit in the middle of the groove so that the cover does not block the positioning
light beam
9.11.6.2 Tilt light
1. Detach the Layer laser cover from the lower shelf by detaching a screw on the
underside and pulling it out.
2. If you need to rotate the light, loosen the screw A. If you need to reposition the light
assembly, loosen the screw B and move the assembly sideways by turning the screw
C.
3. Tighten the screws A and B.
4. Attach the Geometry Calibration Phantom to the device.
5. GUI: Select Device geometry calibration or PAN geometry calibration from
Calibrations menu.
6. All PAN/3D patient positioning lights come on. Check that the Tilt light hits the
vertical groove on the phantom to ensure that the positioning light is correctly
aligned.