KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual page 197

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

Electrical connections
Nominal mains voltage
Input power frequency
Nominal current
Main fuses (F1 & F2)
External warning light fuse (F3)
Power consumption
Maximum impedance of mains
Recommended circuit breaker
Mains cord
Exposure controller cable
Data communication cable
Positioning laser lights
Panoramic laser lights (3)
Cephalostat laser lights (1)
3D laser lights (3)
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
100–120 VAC / 220-240 VAC
Tolerance: ± 10%
50 / 60 Hz
10A @ 220–240 VAC, 15A @ 100 – 120 VAC
220–240Vac:
Littelfuse 215 (Time–Lag) 10A
Cooper Bussman (Time Delay) S505H–10–R
100–120Vac:
Littelfuse 326 (SlowBlow) 15A
Cooper Bussman (Time Delay) MDA–15
Cooper Bussmann (Time Delay) S506–2–R
2A
2.3 kVA @ 220–240 VAC, 1.65 kVA @ 100–
120 VAC
0.2 Ω
max. 30 mA
EU 230V: 16 A / 250 VAC; 50/60 Hz
H05VV–F3G1.5
US 115V: 15 A / 125 VAC 50/60 Hz SJT 3 x
AWG 14
US 230V: 15 A / 250 VAC 50/60 Hz SJT 3 x
AWG 14
Remote exposure switch: DINFLEX–YY 4 x
AWG 26, max. 15m
CAT6 UTP Ethernet cable
IEC 60825–1/2001 (CLASS 1 LASER
PRODUCT)
10 Technical data
197
D519207, 19