9 Installation

9.2.2 Space requirements

The space requirements for the device to be installed and operated depend on the
installation angle of the device. See chapter
device dimensions.
NOTICE! A full scale installation template is supplied with the device to help
selecting the optimal installation position and angle to which the device can be
installed in the selected room.
BEFORE INSTALLING THE DEVICE, ENSURE THAT:
• There is enough space at the front and sides of the device to allow patients to enter
and exit the device easily. Patients in wheelchairs will require more space than
standing patients.
• There is enough space for the operator to position the patient correctly.
• The device is positioned at least 1.5m (59 in) away from any non-medical electrical or
electromechanical equipment.
• The device is positioned so that the operator who is protected from radiation can see
and hear the patient during the imaging, while maintaining visibility to the GUI.
• If the main device is installed in an angle of 15° or more, the plywood supports may
come in the way of the installation and need to be removed. See more details in
chapter
Device dimensions
Recommendations for minimum installation space for a device without CEPH
Depth
Width
Height
Recommendations for minimum installation space for a device with CEPH
Depth
Width
Height
* A minimum measure achievable only in a specific installation position, will require
the device to be rotated during the installation. See chapter
dimensions
on page 206 for more detailed measurements.
84
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 203.
One man installation
180 cm / 70.9 in
110 cm / 43.3 in
210 cm / 81.9 in
One man installation
180 cm / 70.9 in
200 cm / 82.7 in *
210 cm / 81.9 in
Device dimensions
on page 203 for detailed
Two man installation
150 cm / 59 in
150 cm /59 in
Two man installation
150 cm / 59 in *
Cephalometric device
ORTHOPANTOMOGRAPH
OP 3D
D519207, 19