10 Technical data
Table 4: Separation distances
Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC
60601-1-2
The PCX-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The user of PCX-1 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the PCX-1 as
recommended below, according to the maximum output power of the communications equipment.
Frequency of Transmitter
Equation
Rated Maximum Output
Power of Transmitter
(watts)
0,01
0,1
1
10
100
For transmitters rated at maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTICE! At 80 MHz, the separation distance for the higher frequency range applies.
NOTICE! These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
WARNING! *** Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12") to any part of the device,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result. See
USE LIMITATION:
External components
WARNING! Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this device
and result in improper operation.
216
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
150 kHz to 80 MHz
Separation Distance
(meters)
0,20***
0,63
2
6,32
20
Table 4: Separation distances
80 MHz to 800 MHz
Separation Distance
(meters)
0,06***
0,19***
0,6
1,90
6
on page 216.
ORTHOPANTOMOGRAPH
800 MHz to 2.7 GHz
Separation
Distance(meters)
0,06***
0,19***
0,6
1,90
6
OP 3D