7. Re-align the light if it still not aligned correctly.
8. Re-install the detached cover.
9.11.6.3 CEPH FH light
1. Remove the PAN/3D tubehead rear cover by detaching the 3 indicated screws and
pull the cover outwards. The screw on the middle of the cover has a screw cover cap
over it.
2. Remove the PAN/3D tubehead front cover by detaching 4 screws holding the cover
on the tubehead frame and pulling it outwards.
3. The positioning light is held in place in a fixture with one screw. Loosen the screw to
release the light.
4. Re-align the light by rotating it and tilting it up/down.
5. Tighten the screw.
6. GUI: Select CEPH head support calibration from Calibrations menu.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation