PATIENT SIZE SETTINGS:
Adjust the default mA and kV values for medium size patient preset. Other patient size
selections are changed relatively to the adjustment made.
Adjustments made to the 3D imaging program with standard resolution are applied
relatively to other resolution selections as well.
The default values for each program are listed in chapter
values
on page 201.
NOTICE! The available programs and and patient size presets depend on the
device configuration.

9.10.2 Mains power configuration

Detach the indicated screw on the back panel cover and lift it off to access the device
back panel.
9.10.2.1 Mains connection type configuration
IEC 60601-2-63 standard requires extraoral X-ray devices to be installed permanently.
The device can, however, be installed with plug connection if the local regulations allow
this. In these cases, order a mains cord with plug from the manufacturer.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
Patient size setting default
9 Installation
163
D519207, 19