9.6.1 Accessing the calibrations

A complete list of calibration programs, required to be performed after the device
installation, can be unlocked through the Service mode menu on the GUI.
NOTICE! The calibrations programs unlocked through the Service mode menu
are intended for service technicians only, and not to be performed by the end user
of the unit.
NOTICE! The device's country selection needs to be configured correctly before
performing the calibrations. See chapter
1. Attach the device mirror for the duration of the calibrations.
2. Power the device on and establish a connection to the workstation. See chapters
Powering on the device
for more information.
3. Go to device settings on the GUI.
4. Select Service mode menu and enter PIN code 0612.
5. The GUI informs that additional options have been added and goes to Installation
mode.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 46 and
OP 3D
Device settings
on page 158.
Preparing the device for imaging
9 Installation
on page 46