3. Press and hold the exposure button down to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
5. The calibration image appears on the GUI.
NOTICE! The produced calibration image may contain artifacts (e.g. stripes)
at this point.
6. Acknowledge the calibration result by pressing the OK button.

9.6.20 CEPH pixel calibration

This program calibrates the 2D sensor for Cephalometric imaging.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
NOTICE! The PAN/3D Tubehead front cover must be in place before performing
this calibration.
1. Turn the CEPH head support to LAT position.
2. Move the nasion support to its outmost position an turn it down.
3. Select CEPH pixel calibration from the Calibrations menu.
4. Protect yourself from radiation.
5. Press and hold the exposure button down to take the calibration image.
6. When the exposure warning stops and the program end tone is played, the program
is complete.