2. Attach the device column on the exhibition stand. See the installation instructions
supplied with the exhibition stand for more details.
3. If the device is used for other than exhibition use, attach the column to the wall
according to instruction in chapter
4. Complete the installation according to the normal installation instructions in chapter
Carriage installation

9.10.8 Repacking the device

If the device needs to be re-located it must be dismantled and is advisable to be
repacked for safer and easier transportation.
1. GUI: Go to device settings.
2. GUI: Select Service mode menu and enter PIN code 0612.
3. GUI: Select Service menu.
4. Select Drive to packing position.
NOTICE! The device supports driving to Packing position in cases of Z-
movement failure (for service technicians only).
5. Press and hold down the exposure switch to drive the unit to the correct height.
6. Power the device off.
7. Detach the carriage back panel cover.
8. Detach the device from mains power and detach any external cables.
9. Detach the covers in reverse order to what is described in chapter
covers
NOTICE! It's recommended to wrap protective material around the covers
and place them in boxes to protect them against scratches.
10. If the device has a cephalometric unit installed, detach the cephalometric unit covers
and disassemble the cephalometric unit in reverse order to what is described in
chapters
on page 103.
NOTICE! Do not remove any parts that aren't instructed to be attached
during the installation.
11. Detach the column interconnection cables, install the packaging supports back to
their original place and install the plywood supports in reverse order to what is
described in chapter
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 96.
on page 141.
Cephalometric unit covers
Carriage installation
OP 3D
Column installation
on page 87
on page 146 and
Cephalometric unit installation
on page 96.
9 Installation
Main device
171
D519207, 19