KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual page 139

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

4. Pull the nasion support and ear rods to their outmost position and turn them aside.
5. Select CEPH position calibration from the Calibrations menu.
6. Protect yourself from radiation.
7. Press and hold the exposure button down.
8. When the device's movements stop and the program end tone is played, the
program is complete.
9. The calibration image appears on the GUI.
10. Acknowledge the calibration result by pressing the OK button.
11. Remove the CEPH calibration rod from the head support and re-attach the lid.
NOTICE! Repeat the CEPH geometry calibration if the GUI requests it.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
139
D519207, 19

Advertisement

Table of Contents
loading

Table of Contents