4. Pull the nasion support and ear rods to their outmost position and turn them aside.
5. Select CEPH position calibration from the Calibrations menu.
6. Protect yourself from radiation.
7. Press and hold the exposure button down.
8. When the device's movements stop and the program end tone is played, the
program is complete.
9. The calibration image appears on the GUI.
10. Acknowledge the calibration result by pressing the OK button.
11. Remove the CEPH calibration rod from the head support and re-attach the lid.
NOTICE! Repeat the CEPH geometry calibration if the GUI requests it.