9.5 Communication link configuration

1. Install the Dental Imaging software to the workstation.
2. Connect the device to the local area network (LAN) or directly to the workstation
with an Ethernet cable. See the following chapters for more detailed information.
NOTICE! The unit requires 1Gb Ethernet connection between the unit and the
reconstruction workstation. The workstation must be connected to the unit in
accordance with IEC 60601-1-1.
3. Connect the device to mains power.
4. Power the device on. Power switch is located at the rear of the carriage.
5. Configure the imaging software to connect to the device. See software instructions
for more information.
6. When the communication link is established, the device GUI appears on the screen.
7. Complete the device initialization by pressing the HOME button on the patient
positioning panel.

9.5.1 Network Configuration

In network configuration, the unit is connected to the workstation via a 1Gb network
switch. With this configuration, the PC is connected to the unit and to other devices
in the network with a single network card. The unit is also accessible from other
workstations in the same network (same subnet).
By default, the device is configured to retrieve its IP address automatically from the
DHCP server in the network.
For the network configuration with static IP address, you may need a dedicated IP-
address for the unit to be reserved by your local system administrator.
When configuring the IP address or DHCP server, the device can have any IP-address,
but it and the reconstruction workstation must be in the same subnet. Subnet mask is
recommended to be set as 255.255.255.0.
NOTICE! The device does not support connection to IPv6 or L3 (routable)
networks. Only IPv4 is supported.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
115
D519207, 19