NOTICE! You should also compare the new QC image to the reference image
taken during the installation or the latest service. Doing this helps you ensure that
the image quality has remained constant.
NOTICE! The line pair resolution depends also on the other factors than the
device itself, for example the imaging Software configurations. According to the
standards, the distinguishable line pair resolution must be 2.5 LP/mm or better.
13. If the image fails on any of the previously listed criteria, redo the QC program. If it
fails again, re-run the CEPH calibrations according to
69 or contact service.
14. Detach the CEPH QC phantom holder and 2D QC test phantom from the device.
15. Detach the copper filter from the CEPH tubehead if it was used.