6 Using the device

6.2 Powering on the device

1. Power the device on. Power switch is located at the rear of the carriage.
2. The device starts initialization.
3. Complete the device initialization by pressing the HOME button on the patient
positioning panel when indicator light starts blinking in blue. The device GUI also
indicates when HOME button can be pressed.
NOTICE! The device moves up or down when HOME button is pressed.
NOTICE! The device enters a power save mode after 30 minutes of inactivity.
While in the power save mode, the device cooling fan is not active, but the
patient positioning panel lights are lit. The device wakes up from the power
save mode when connection to it is established or when the HOME button is
pressed.

6.3 Preparing the device for imaging

1. Power up the workstation.
2. Launch the dental imaging software.
3. Imaging SW: Select the patient for examination.
4. Imaging SW: Select the device to establish a connection.
NOTICE! Refer to the dental imaging software documentation for more
details.
5. The device GUI opens on the workstation when the connection to the device is
established.
NOTICE! The device must be powered on and initialized before the
connection can be established.